How to adjust calibration on Lumity Cooking Oil Tester
Sally Spears
Last Update il y a 2 ans

Once the calibration has been checked and the deviation is greater the 2% the calibration needs to be adjusted on the Cooking Oil Tester
If you have not checked the calibration you need to complete this step to know how much to adjust the value on the Cooking Oil Tester
If you are unsure how to check the calibration please see Checking Calibration of the Cooking Oil Tester document in support.squizify.com
Press the H key and hold for 3 seconds to go to the menu settings

In the menu settings, move up or down using the up and down arrows, to select TPM Adjust
Press the H key to enter TPM Adjust area
Using the up and down keys input the adjustment value.
- You will know how much to adjust this value once you have completed the Checking Calibration of the Cooking Oil Tester
TPM adjustment value = Standard value - measurement value
Then press H to save and exit the menu
Move to return and press H to return to the measurement screen
Your calibration has been adjusted successfully!