How to check the calibration of the Lumity Cooking Oil Tester (TPM)
Sally Spears
Last Update 2 jaar geleden

To ensure the Cooking Oil Tester produces accurate readings the calibration must be checked regularly.
The calibration can be adjusted on the probe itself
To calibrate the Cooking Oil Tester you mush have:
- A Lumity TPM Cooking Oil Tester
- Access to hot water
- Emerson Calibration Test Oil (Check that it is in date!)
If you do not have the test oil contact Squizify to order this product
Clean the Sensor using a Cloth or Disposable Paper Towel

Place the reference oil container in a jug of hot water
Warm the oil until it reaches 50°C

Immerse the Cooking Oil Tester into reference oil and take a reading.

Compare reading with the label value.
- If deviation is more than 2%, adjustment is necessary.
- If deviation is less that 2% or at 2% no adjustment is needed
You have checked the Calibration of the Lumity Cooking Oil Tester!
If your Lumity Cooking Oil Tester needs to be adjusted please see "How to adjust calibration on the Lumity Cooking Oil Tester" in the Support Centre.